General MRI compatibility conditions - Before every MRI, the device must be checked and correctly programmed. - The manufacturer’s warranty applies to the system composed of the device and lead (by the same manufacturer). - Should there be any restriction difference between the lead and the device, the most restrictive applies to the whole device + lead system (e.g. in case of a 1.5T device and a 3T lead, the restriction for the entire device + lead system is 1.5T). - In case of an abandoned lead or epicardial lead, the system is not MRI-compatible.
In combination with a Kora 250 pacemaker. Please refer to MRI Solutions KORA 250 Manual (cardiologist and radiologist).
Last update
Monday, 29 April 2024
BEFLEX (RF45D, RF46D)
Field strength
1,5 T with exclusion
Exclusion zone
Thorax
Specific conditions
In combination with a Kora 100 pacemaker. Please refer to MRI Solutions KORA 100 Manual (cardiologist and radiologist).
Last update
Monday, 29 April 2024
BEFLEX (RF45D, RF46D)
Field strength
3T full body
Specific conditions
In combination with MicroPort CRM MRI Conditional devices . Please refer to MRI Solutions Manual of the implanted device (cardiologist and radiologist).
Last update
Monday, 29 April 2024
VEGA (R45, R52, R58)
Field strength
1,5T full body
Specific conditions
In combination with a Kora 250 pacemaker. Please refer to MRI Solutions KORA 250 Manual (cardiologist and radiologist).
Last update
Monday, 29 April 2024
VEGA (R45, R52, R58)
Field strength
1,5 T with exclusion
Exclusion zone
Thorax
Specific conditions
In combination with a Kora 100 pacemaker. Please refer to MRI Solutions KORA 100 Manual (cardiologist and radiologist).
Last update
Monday, 29 April 2024
VEGA (R45, R52, R58)
Field strength
3T full body
Exclusion zone
No exclusion zone
Specific conditions
In combination with ENO/TEO/OTO and ALIZEA/BOREA/CELEA pacemakers. Please refer to MRI Solutions ENO/TEO/OTO Manual or MRI Solutions ALIZEA/BOREA/CELEA Manual (cardiologist and radiologist)
Last update
Monday, 29 April 2024
XFine (JX24D, JX25D, TX25D, TX26D)
Field strength
1,5 T with exclusion
Exclusion zone
Thorax
Specific conditions
In combination with a Kora 100 pacemaker. Please refer to MRI Solutions KORA 100 Manual (cardiologist and radiologist).