General MRI compatibility conditions - Before every MRI, the device must be checked and correctly programmed. - The manufacturer’s warranty applies to the system composed of the device and lead (by the same manufacturer). - Should there be any restriction difference between the lead and the device, the most restrictive applies to the whole device + lead system (e.g. in case of a 1.5T device and a 3T lead, the restriction for the entire device + lead system is 1.5T). - In case of an abandoned lead or epicardial lead, the system is not MRI-compatible.
Specific conditions from the manufacturer
Download pdf (document provided by the manufacturer)
According to Abbott compatible MRI cans
1,5T: in combination with Ellipse CD1377-36Q, CD1377-36QC, CD2377-36Q, CD2377-36QC or Fortify Assura CD1359-40Q, CD1359-40QC, CD2359-36Q, CD2359-40QC or Quadra Assura MP CD3371-40Q, CD3371-40QC or Quadra Assura CD3367-40Q, CD3367-40QC; or Gallant CDVRA500Q, CDDRA500Q, CDHFA500Q; and compatible pacing leads.
3 T: in combination with CDVRA500Q, CDDRA500Q, CDHFA500Q and compatible pacing leads.
Last update
Saturday, 24 February 2024
Optisure - LDA210Q, LDA220Q (longueurs 58, 65cm)
Field strength
1,5T full body 3T full body
Specific conditions
According to Abbott compatible MRI cans :
1,5T full body : Optisure - LDA210Q, LDA220Q (lengths 58, 65cm) in combination with Ellipse CD1377-36Q, CD1377-36QC, CD2377-36Q, CD2377-36QC or Fortify Assura CD1359-40Q, CD1359-40QC, CD2359-36Q, CD2359-40QC or Quadra Assura MP CD3371-40Q, CD3371-40QC or Quadra Assura CD3367-40Q, CD3367-40QC
1,5T et 3T full body: Optisure - LDA210Q, LDA220Q (lengths 58, 65cm) in combination with Gallant CDVRA500Q, CDDRA500Q and CDHFA500Q.