General MRI compatibility conditions - Before every MRI, the device must be checked and correctly programmed. - The manufacturer’s warranty applies to the system composed of the device and lead (by the same manufacturer). - Should there be any restriction difference between the lead and the device, the most restrictive applies to the whole device + lead system (e.g. in case of a 1.5T device and a 3T lead, the restriction for the entire device + lead system is 1.5T). - In case of an abandoned lead or epicardial lead, the system is not MRI-compatible.
Specific conditions from the manufacturer
Download pdf (document provided by the manufacturer)
EDORA 8 pacemaker associated to MRI compatible BIOTRONIK leads
EDORA 8 SR-T/DR-T: 1.5T FBS + 3T with Thoracic Exclusion Zone in combination with Solia S45 lead
EDORA 8 HF-T/HF-T QP: 1.5 FBS in combination with Solia S 45 lead
Last update
Thursday, 07 March 2024
ELUNA 8 (SR-T, DR-T) [394971, 394969]
Field strength
1,5T full body 3T with exclusion
Exclusion zone
Thorax for the 3T
Specific conditions
ELUNA 8 pacemaker associated to MRI compatible BIOTRONIK leads
Last update
Thursday, 07 March 2024
ELUNA 8 HF-T [394968]
Field strength
1,5 T with exclusion
Exclusion zone
Thorax
Specific conditions
ELUNA 8 pacemaker associated to MRI compatible BIOTRONIK leads
Last update
Thursday, 07 March 2024
EVIA (SR-T, DR-T) [371998, 371996]
Field strength
1,5T full body 3T with exclusion
Exclusion zone
Thorax for the 3T
Specific conditions
EVIA pacemaker associated to MRI compatible BIOTRONIK leads
Last update
Thursday, 07 March 2024
EVIA HF-T [381535]
Field strength
1,5 T with exclusion
Exclusion zone
Thorax
Specific conditions
EVIA pacemaker associated to MRI compatible BIOTRONIK leads